Location
- Research Support Building (CACTUS)
- Rúa de Constantino Candeira, 1. Campus Vida , 15782Santiago de Compostela
- Phones
- 881 816 242
-
Email
- esteban.guitian [at] usc.es
- Schedule
- Monday to Friday, from 10 a.m. to 2 p.m. and from 4 p.m. to 6 p.m.
Mass spectrometer with 122 cm linear TOF (Time of Flying) analyser.
- Ionisation source: MALDI.
- Detection system: with MCP (Micro Chanel Plate) of very high sensitivity and high dynamic range.
- Power supply: high voltage +25/-20 kV for ion acceleration.
- Laser N2 type LTB MNL 106.
- SCOUT source, with microtiter plate format. Sample holder with observation optics.
- AnchorChip sample concentration system to increase the equipment's own sensitivity by 10 to 100 times.
- Electrostatic ion reflector for higher resolution and accuracy in pulsed ion extraction (PIE) mode and a second special high-sensitivity MCP type detector.
- FAST Module (Fragmentation Analysis and Structural TOF-MS). Accessory for PSD (Post-Source Decay) experiments.
Mass Spectrometry and Proteomics Unit
- Research Support Building (CACTUS)
- Rúa de Constantino Candeira, 1. Campus Vida , 15782Santiago de Compostela